Pressure Decay Leak Testing of Valves

Pressure Decay Leak Testing of Valves
Pressure Decay Leak Testing of Valves
A pressure decay leak test is a test which observes a drop in pressure over a specified time to determine if a leak is present. The main advantage of the pressure decay leak test is that it is an objective test. An operator connects the test part to the pressure decay leak tester, presses the start button, and waits for the test to complete and output the test results.

The way a pressure decay leak test works is by filling the test part with pressurized air. After the pressure setpoint has reached, the test valve of the pressure decay leak tester will close. The setpoint pressure is determined by the part. This is something that has to be figured out through some initial testing. Too little pressure and the test results may not be as sensitive because pressure is correlated to detectability, too much pressure and you risk damaging the test part. In our experience, we have seen pressures as little as 1 psi to as high as 250psi. It is recommended to start with lower pressures and work your way up to higher pressure before this test procedure is standardized and moved onto the production or quality assurance. After the test valve has closed, the air inside the test part will settle. The Pressure Decay Leak Testing Instrument will continuously monitor the pressure inside the part during the whole test. If a leak is present, the Pressure Decay Instrument will detect this in the form of decrease in pressure. If the pressure inside the test part falls below the pressure setpoint, the result will be a “FAIL” and output on the HMI Screen. If on the other hand, the pressure stays above the pressure setpoint, the results will be a “PASS” and shown on the HMI Screen. How much of a pressure drop is considered a passing part must also be determined prior to standardizing this test.

A valve manufacturer reached out to us because they needed an objective test to test the closure quality of their valves. Their test was simple, they needed to close the valve and determine if the pressure drop across the valve was zero over the course of 3 minutes. Their acceptable pressure drop was less than 0.1 psi at a pressure of 125 psi. We made them a test jig where they placed the valve into the test fixture, clamped it closed, pressed the START button, and in 3 minutes they had an answer. In order to increase the test sensitivity, and to give us the best signal-to-noise ratio, we had to make sure that the testing volume of the valve was as small as possible. This was done be decreasing the distance of the airline and by directly connecting to the valve which minimized all the head space of the air testing line. If you are looking to perform a pressure decay leak test. Contact us to see how we can help.

Some of our Popular Items

Our clients prefer to work with us because we are Experts in Custom Fabrication (especially Polymer Fabrication). You should check out some of our other items we carry; click on the links below.

Vacuum Oven Systems
Vacuum Ovens are vacuum chambers and ovens combined in one system. Vacuum ovens are necessary during processes which require the specimen to be subjected to a hypobaric and hyperthermal environment. Our Vacuum Oven Systems enable to you run a complete vacuum vs time and temperature vs time profile. We also carry portable vacuum oven systems and custom vacuum oven systems.
Wire Vacuum Feedthroughs
Wire Vacuum Feedthroughs are wires that are run from the outside of the chamber towards the inside of the vacuum chamber. We carry a wide variety of wires and wire gauges that can be used with any vacuum chamber.
Our Work: Submersion Leak Testing of Explosive or Volatile Products
Ever wondered how to leak test packaged volatile products? Do you know that there are countless unstable products out there that have to be leak tested as part of quality control? By volatile we mean products which ignite, explode, shatter, melt, or are hazardous. Keep in mind that hazardous material for whatever reason has be undergo quality control as well.
Related Articles: Selecting a Proper Vacuum Pump for your Acrylic Vacuum Chamber
In this article we will talk about Vacuum Pumps - but also about other equipment which is used to produce vacuum inside a closed system or chamber. We will talk about the workings of a vacuum pump, vacuum generation, pump down, pump horsepower. This article is intended to provide you a practical guide which should help you select the appropriate vacuum pump for your system.